Document Type
Article
Publication Date
1-1-2017
Identifier
PMCID: PMC5621639 DOI: 10.1016/j.jspd.2016.09.003
Abstract
BACKGROUND: Distraction-type rods mechanically stabilize the thorax and improve lung growth and function by applying distraction forces at the rib, spine, pelvis, or a combination of locations. However, the amount of stability the rods provide and the amount the thorax needs is unknown.
METHODS: Five freshly frozen and thawed cadaveric thoracic spine specimens were tested for lateral bending, flexion/extension, and axial rotation in displacement control (1°/sec) to a load limit of ±5 Nm for five cycles after which a growth-friendly unilateral rod was placed in a simulated rib-to-lumbar attachment along the right side. The specimens were tested again in the same modes of bending. From the seven Optotrak Orthopedic Research Pin markers (Northern Digital Inc., Waterloo, Ontario, Canada) inserted into the top potting to denote T1, and the right pedicles at T2, T4, T5, T8, T9, and T11 and the Standard Needle Tip Pressure Transducers (Gaeltech, Isle of Skye, Scotland) inserted into the T4/T5 and T8/T9 discs, motion, stiffness, and pressure data were calculated. Parameters from the third cycle of the intact case and the construct case were compared using two-tailed paired t tests with 0.05 as the level of significance.
RESULTS: With the construct attached, the T1-T4 segment showed a 30% increase in neutral zone stiffness during extension (p = .001); the T8-T12 segment experienced a 63% reduction in the in-plane range of motion during flexion (p = .04); and the T8/T9 spinal motion unit had a significant decrease of 24% in elastic zone stiffness during left axial rotation (p = .04).
CONCLUSIONS: It is clear the device as tested here does not produce large biomechanical changes, but the balance between providing desired changes while preventing complications remains difficult.
Journal Title
Spine Deform
Volume
5
Issue
1
First Page
11
Last Page
17
MeSH Keywords
Thoracic Vertebrae; Scoliosis; Internal Fixators; Biomechanical Phenomena; Adolescent/growth and development
Keywords
Autofusion; Biomechanics; Intradiscal pressure; Scoliosis
Recommended Citation
Galvis S, Arnold J, Mannen E, et al. Biomechanical Evaluation of a Growth-Friendly Rod Construct. Spine Deform. 2017;5(1):11-17. doi:10.1016/j.jspd.2016.09.003
Included in
Musculoskeletal System Commons, Orthopedics Commons, Pediatrics Commons, Surgery Commons, Surgical Procedures, Operative Commons