Comparison of three multiplex gastrointestinal platforms for the detection of gastroenteritis viruses
Document Type
Article
Publication Date
10-2017
Identifier
DOI: 10.1016/j.jcv.2017.08.012; PMCID: PMC5940486
Abstract
Background Viruses are major etiological agents of childhood gastroenteritis. In recent years, several molecular platforms for the detection of viral enteric pathogens have become available.
Objective/study design We evaluated the performance of three multiplex platforms including Biofire's Gastrointestinal Panel (FilmArray), Luminex xTAG® Gastrointestinal Pathogen Panel (GPP), and the TaqMan Array Card (TAC) for the detection of five gastroenteritis viruses using a coded panel of 300 archived stool samples.
Results The FilmArray detected a virus in 199 (96.1%) and the TAC in 172 (83.1%) of the 207 samples (187 samples positive for a single virus and 20 samples positive for more than one virus) whereas the GPP detected a virus in 100 (78.7%) of the 127 (97 positive for one virus and three positive for more than one virus) samples. Overall the clinical accuracy was highest for the FilmArray (98%) followed by TAC (97.2%) and GPP (96.9%). The sensitivity of the FilmArray, GPP and TAC platforms was highest for rotavirus (100%, 95.8%, and 89.6%, respectively) and lowest for adenovirus type 40/41 (97.4%, 57.9% and 68.4%). The specificity of the three platforms ranged from 95.6% (rotavirus) to 99.6% (norovirus/sapovirus) for the FilmArray, 99.6% (norovirus) to 100% (rotavirus/adenovirus) for GPP, and 98.9% (astrovirus) to 100% (rotavirus/sapovirus) for TAC.
Conclusion The FilmArray demonstrated the best analytical performance followed by TAC. In recent years, the availability of multi-enteric molecular testing platforms has increased significantly and our data highlight the strengths and weaknesses of these platforms.
Journal Title
Journal of Clinical Virology
Volume
95
First Page
66
Last Page
71
Keywords
Adenovirus type 40/41, Astrovirus, Gastroenteritis viruses, Norovirus, Rotavirus, Sapovirus
Recommended Citation
Chhabra P, Gregoricus N, Weinberg GA, et al. Comparison of three multiplex gastrointestinal platforms for the detection of gastroenteritis viruses. J Clin Virol. 2017;95:66-71. doi:10.1016/j.jcv.2017.08.012
Comments
Grant support