Positioning of Diaphragm Electrical Activity Catheter in Premature Infants.

Document Type

Article

Publication Date

7-2026

Identifier

DOI: 10.1177/19433654261424880

Abstract

BACKGROUND: Neurally-adjusted ventilatory assist (NAVA) is a mode of ventilation that uses the electrical activity of the diaphragm (EAdi) to synchronize and to control delivered pressure. EAdi is measured with a naso-/orogastric feeding tube, containing an array of sensors. Proper placement of the EAdi catheter is critical for optimal synchronization and support. Positioning the catheter requires (1) predicting the insertion distance and (2) verification using a dedicated positioning window on the ventilator, based on electrocardiogram (ECG) and EAdi signals. In preterm infants, no study has shown that the ventilator method results in proper positioning of the feeding tube tip in the stomach. We aimed to evaluate the position of the EAdi sensors (by the catheter positioning window) with the feeding tube tip position (measured by radiographs).

METHODS: This was a multi-center (5 sites), prospective study. Eligibility: infants with weight 400-2,000 g and either had or were planning to have an EAdi catheter placed. Screenshots of the positioning window and radiographs were taken.

RESULTS: Sixty-seven infants were included. Median study weight was 1,250 g (interquartile range 1,026-1,448 g). The ventilator's positioning window revealed that all insertions were suitable with respect to the diaphragm/sensors position. Radiographs indicated 92% of insertions had the tip of the catheter appropriately in the body of the stomach. In 5 infants, the catheter tip was either touching the greater curvature of the stomach or near the pylorus.

CONCLUSIONS: In our cohort, EAdi catheter insertion using guidance from the ECG/EAdi signals of the electrode array provided a safe method for tube positioning, with regard to both enteral feeding and obtaining appropriate EAdi signals for optimal ventilator support.

Journal Title

Respiratory care

Volume

71

Issue

7

First Page

740

Last Page

747

MeSH Keywords

Humans; Diaphragm; Infant, Newborn; Infant, Premature; Prospective Studies; Female; Interactive Ventilatory Support; Male; Electrocardiography; Enteral Nutrition

PubMed ID

41761951

Keywords

diaphragm electrical activity; feeding tube positioning; neurally adjusted ventilatory assist