Document Type

Article

Publication Date

8-1-2012

Identifier

PMCID: PMC3568492 DOI: 10.1016/j.echo.2012.05.004

Abstract

BACKGROUND: Clinical trials often rely on echocardiographic measures of left ventricular size and function as surrogate end points. However, the quantitative impact of factors that affect the reproducibility of these measures is unknown. To address this issue, the National Heart, Lung, and Blood Institute-funded Pediatric Heart Network designed a longitudinal observational study of children with known or suspected dilated cardiomyopathy aged 0 to 22 years from eight pediatric clinical centers.

METHODS: Clinical data were collected together with 150 echocardiographic indices of left ventricular size and function. Separate observers performed duplicate echocardiographic imaging. Multiple observers performed measurements from three cardiac cycles to enable assessment of intraobserver and interobserver variability. The impacts of beat averaging (BA), observer type (local vs core), and variable type (areas, calculations, dimensions, slopes, time intervals, and velocities) on measurement reproducibility were studied. The outcome measure was percentage error (100 × difference/mean).

RESULTS: Of 173 enrolled subjects, 131 met criteria for dilated cardiomyopathy. BA, variable type and observer type all influenced percentage error (P < .0001). Core interobserver percentage error (medians, 11.4%, 10.2%, and 9.3% for BA using one, two, and three beats, respectively) was approximately twice the intraobserver percentage error (medians, 6.3%, 4.9%, and 4.2% for BA using one, two, and three beats, respectively). Slopes and calculated variables exhibited high percentage error despite BA. Chamber dimensions, areas, velocities, and time intervals exhibited low percentage error.

CONCLUSIONS: This comprehensive evaluation of quantitative echocardiographic methods will provide a valuable resource for the design of future pediatric studies. BA and a single core lab observer improve the reproducibility of echocardiographic measurements in children with dilated cardiomyopathy. Certain measurements are highly reproducible, while others, despite BA, are poorly reproducible.

Journal Title

Journal of the American Society of Echocardiography : official publication of the American Society of Echocardiography

Volume

25

Issue

8

First Page

842

Last Page

854

MeSH Keywords

Adolescent; Adult; Cardiomyopathy, Dilated; Child; Child, Preschool; Comorbidity; Echocardiography; Heart Ventricles; Humans; Image Enhancement; Image Interpretation, Computer-Assisted; Infant; Infant, Newborn; Male; Organ Size; Prevalence; Reproducibility of Results; Risk Factors; Sensitivity and Specificity; United States; Ventricular Dysfunction, Left; Young Adult